Agilent Technologies A.18.00 Manual de usuario Pagina 200

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 581
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 199
200
S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\efpkc.sec
Equivalent Front-Panel Key Commands
INSTRUMENT STATE Keys
ADRS
The ADRS key displays the Test Set’s GPIB address in the upper left-hand corner
of the CRT. There is no equivalent GPIB command for the ADRS key. The current
address can also be viewed by looking at the
HP-IB Adrs field on the I/O
CONFIGURE screen.
The Test Set’s GPIB address can be changed through GPIB by using the
:CONFigure:BADDress commands. If the Test Set’s GPIB address is changed
programmatically, all future GPIB commands must use the new address.
Syntax
CONFigure:BADDress <integer number>
Example
OUTPUT 714;"CONF:BADD 15"
This sets the Test Set’s GPIB address to 15.
The Test Set’s GPIB address can be queried through GPIB by using the
:CONFigure:BADDress? commands.
Syntax
‘CONFigure:BADDress?
Example
OUTPUT 714;"CONF:BADD?"
ENTER 714;Address
This queries the Test Set’s GPIB address.
Vista de pagina 199
1 2 ... 195 196 197 198 199 200 201 202 203 204 205 ... 580 581

Comentarios a estos manuales

Sin comentarios