519
Chapter 8, Programming the Call Processing Subsystem
Programming the CALL CONFIGURE Screen
Detector
This field is used to select the type of detector used to measure the amplitude of
the audio signal being analyzed on the
ANALOG MEAS screen. The Detector
field is imported from the
AF ANALYZER screen and is programmed exactly as it
is on its home screen.
See “AF Analyzer” on page 97 for programming command
syntax.
TX Pwr Zero
The TX Pwr Zero function establishes a 0.0000 W reference for measuring RF
power at the RF IN/OUT port. The
TX Pwr Zero field is imported from the RF
ANALYZER
screen and is programmed exactly as it is on its home screen. See “RF
Analyzer” on page 161
for programming command syntax.
CAUTION: RF power must not be applied while zeroing the power meter.