S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb
Chapter 5, Advanced Operations
Status Reporting
Reading the Event Register
Syntax
STATus:COMMunicate:EVENt?
Example
OUTPUT 714;"STAT:COMM:EVEN?"
ENTER 714;Register_value
Clearing the Event Register
The EVENT register is cleared whenever it is queried or whenever the Common
Command *CLS is sent to the Test Set.
Reading the Enable Register
Syntax
STATus:COMMunicate:ENABle?
Example
OUTPUT 714;"STAT:COMM:ENAB?"
ENTER 714;Register_value
Writing the Enable Register
Syntax
STATus:COMMunicate:ENABle <integer>
Example
OUTPUT 714;"STAT:COMM:ENAB 256"
Clearing the Enable Register
The ENABLE register is cleared by writing to it with an integer value of zero.