Agilent B1500A User’s Guide, Edition 13 4- 37
Specifications
Hardware
WGFMU supplemental data
•RSU SMU path
Leakage current: < 100 pA
Residual resistance: < 300 m
Jitter: < 1 ns
Skew between channels: < 3 ns, under no electrostatic discharge condition
Trigger output skew: < 3 ns
Current range change time: < 150 s
This is the time until the measured current settles within ± 0.3 % of the final
result value after the range change.
Minimum timing parameters for current measurement
: See Table 4-22.
Minimum timing parameters for voltage measurement
: See Table 4-23.
Table 4-17 Measurement functions, voltage forcing, voltage measurement, and current
measurement
Mode Function V force ranges
V measure
ranges
I measure
ranges
Fast IV V force/
I measure,
V
force/
V measure
-3 V to +3 V,
-5 V to +5 V,
-10 V to 0 V,
0
V to +10 V
-5 V to +5 V,
-10
V to
+10
V
1 A, 10 A,
100
A,
1
mA, 10 mA
PG V force/
V measure
-3 V to +3 V,
-5 V to +5 V
-5 V to +5 V -
DC V force/
I measure,
V
force/
V measure
-3 V to +3 V,
-5 V to +5 V,
-10 V to 0 V,
0
V to +10 V
-5 V to +5 V,
-10
V to
+10
V
1 A, 10 A,
100
A,
1
mA, 10 mA
SMU pass
through
Measurement
using SMU
Maximum
±
25 V
- Maximum
±
100 mA
Vista de pagina 176