Vista de pagina 190
Agilent B1500A User’s Guide, Edition 13 4- 51
Specifications
General Specifications
General Specifications
Temperature range
Operating: +5 C to +40 C
Storage: -20 C to +60 C
•Humidity
Operating: 20 % to 70 %RH, non-condensing
Storage: 10 % to 90 %RH, non-condensing
Altitude
Operating: 0 to 2,000 m (6,561 ft.)
Storage: 0 to 4,600 m (15,092 ft.)
Power requirement: 100-240 V (±10 %), 50/60 Hz
Maximum volt-amps (VA): 900 VA
Regulatory compliance
•EMC
IEC 61326-1 / EN 61326-1
ICES/NMB-001
AS/NZS CISPR 11
KC: RRA Notification amending Radio Waves Act Article 58-2
Safety
IEC 61010-1 / EN 61010-1
CAN/CSA-C22.2 No. 61010-1
•Dimensions
B1500A: 420 mm (W) 330 mm (H) 575 mm (D)
B1531A RSU: 45.2 mm (W) 70 mm (H) 82 mm (D)
E5288A ASU: 132 mm (W) 88.5 mm (H) 50 mm (D)
N1255A Connection box: 184.4 mm (W) 61.6 mm (H) 169.6 mm (D)
N1301A-100 SCUU: 148 mm (W) 75 mm (H) 70 mm (D)
N1301A-200 GSWU: 33.2 mm (W) 41.5 mm (H) 32.8 mm (D)