Agilent B1500A User’s Guide, Edition 13 4- 43
Specifications
EasyEXPERT Software
Oscilloscope view
Available for the tracer test using MCSMU modules. The oscilloscope view displays
MCSMU current or voltage measurement data versus time. The pulsed
measurement waveforms appear in a separate window for easy verification of the
measurement timings. This function is useful for verifying waveform timings and
debugging pulsed measurements. It is available when a tracer test has one or more
MCSMU channels being used in pulsed mode. The oscilloscope view can display
the pulse waveform at any (user specified) sweep step of the sweep output.
• Sampling interval: 2 s
• Sampling points: 2000 Sa
• Sampling duration: 22 s to 24 ms
• Marker function
Data read-out for each channel
Resolution: 2 s
• Data saving
Numeric: TXT/CSV/XMLSS
Image: EMF/BMP/JPG/PNG
Category Test items
CMOS Id-Vg, Id-Vd, Vth, breakdown, capacitance, QSCV, etc.
BJT Ic-Vc, diode, Gummel plot, breakdown, hfe, capacitance, etc.
Discrete Id-Vg, Id-Vd, Ic-Vc, diode, etc.
Memory Vth, capacitance, endurance test, etc.
PwrDevice Pulsed Id-Vg, pulsed Id-Vd, breakdown, etc.
NanoTech Resistance, Id-Vg, Id-Vd, Ic-Vc, etc.
Reliability NBTI/PBTI, charge pumping, electromigration, hot carrier
injection, J-Ramp, TDDB, etc.