99
Table 33. Uncorrected System Performance
a
Specifications apply over environmental temperature of 23° ±3 °C, with < 1 °C deviation from the calibration temperature
Description Specification Supplemental Information
Directivity
10 MHz to 45 MHz
b
-- 11 dB (typical)
45 MHz to 2 GHz 24 dB --
2 GHz to 10 GHz 22 dB --
10 GHz to 20 GHz 16 dB --
20 GHz to 40 GHz 16 dB --
40 GHz to 45 GHz 15 dB --
45 GHz to 50 GHz 13 dB --
Source Match - Standard
10 MHz to 45 MHz
b
-- 11 dB (typical)
45 MHz to 2 GHz 23 dB --
2 GHz to 10 GHz 16 dB --
10 GHz to 20 GHz 14 dB --
20 GHz to 40 GHz 10 dB --
40 GHz to 45 GHz 9 dB --
45 GHz to 50 GHz 7.5 dB --
Source Match - Opt UNL, 014 or 014/UNL
10 MHz to 45 MHz
b
-- 11 dB (typical)
45 MHz to 2 GHz 18 dB --
2 GHz to 10 GHz 14 dB --
10 GHz to 20 GHz 12 dB --
20 GHz to 40 GHz 9 dB --
40 GHz to 45 GHz 8 dB --
45 GHz to 50 GHz 6 dB --
Load Match - Standard
10 MHz to 45 MHz
b
-- 11 dB (typical)
45 MHz to 2 GHz 23 dB --
2 GHz to 10 GHz 14 dB --
10 GHz to 20 GHz 10 dB --
20 GHz to 40 GHz 9 dB --
40 GHz to 45 GHz 9 dB --
45 GHz to 50 GHz 8 dB --
Load Match - Opt UNL, 014 or 014/UNL
10 MHz to 45 MHz
b
-- --
45 MHz to 2 GHz 17 dB --
2 GHz to 10 GHz 13 dB --
10 GHz to 20 GHz 10 dB --
20 GHz to 40 GHz 9 dB --
40 GHz to 45 GHz 9 dB --
45 GHz to 50 GHz 7 dB --