Agilent Technologies E8364C Manual de usuario Pagina 61

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 145
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 60
57
Table 19. 85050D Calibration Kit
Standard Configuration and Standard Power Range
(E836xB/C)
Applies to the, E836xB/C analyzers, 85050D (7mm) calibration kit, 85132F flexible test port cable set, and a full 2-port calibration. Also
applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB) Description
0.045 to
2 GHz
2 to
18 GHz
Directivity 40 40
Source Match 39 35
Load Match 40 37
Reflection Tracking
±0.010
+0.02/°C
±0.100
+0.02/°C
Transmission Tracking
±0.013
+0.02/°C
±0.072
+0.02/°C
Vista de pagina 60
1 2 ... 56 57 58 59 60 61 62 63 64 65 66 ... 144 145

Comentarios a estos manuales

Sin comentarios