Agilent Technologies E8364C Manual de usuario Pagina 131

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 145
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 130
127
E8363B/C:
10 Hz IF Bandwidth
45 MHz to 500 MHz < -127 dBm --
500 MHz to 2 GHz < -133 dBm --
2 GHz to 10 GHz < -132 dBm --
10 GHz to 20 GHz < -134 dBm --
20 GHZ to 40 GHz < -125 dBm --
1 kHz IF Bandwidth
45 MHz to 500 MHz < -107 dBm --
500 MHz to 2 GHz < -113 dBm --
2 GHz to 10 GHz < -112 dBm --
10 GHz to 20 GHz < -114 dBm --
20 GHZ to 40 GHz < -105 dBm --
E8364B/C:
10 Hz IF Bandwidth
--
45 MHz to 500 MHz < - 127 dBm --
500 MHz to 2 GHz < - 133 dBm --
2 GHz to 10 GHz < - 132 dBm --
10 GHz to 20 GHz < - 134 dBm --
20 GHZ to 40 GHz < - 125 dBm --
40 GHZ to 50 GHz < - 123 dBm --
1 kHz IF Bandwidth
45 MHz to 500 MHz < -107 dBm --
500 MHz to 2 GHz < -113 dBm --
2 GHz to 10 GHz < -112 dBm --
10 GHz to 20 GHz < -114 dBm --
20 GHZ to 40 GHz < -105 dBm --
40 GHZ to 50 GHz < -103 dBm --
Damage Level
E8362B/C -- + 15 dBm
E8363B/C -- + 15 dBm
E8364B/C -- + 15 dBm
Maximum DC Level
E8362B/C -- + 15 V
E8363B/C -- + 15 V
E8364B/C -- + 15 V