Agilent Technologies E8364C Manual de usuario Pagina 79

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 145
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 78
75
E8363B/C AND E8364B/C Corrected System Performance with WR-28 Connectors
Table 25. R11644A Calibration Kit
Standard Configuration and Standard Power Range
(E8363B/C AND E8364B/C)
Applies to the, E8363B/C AND E8364B/C analyzers, R11644A (WR-28) calibration kit, 85133F flexible test port cable set with the R281A
and R281B launch sets, and a full 2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)
Description
26.5 to
40 GHz
Directivity 50
Source Match 50
Load Match 50
Reflection Tracking
±0.000
+0.03/°C
Transmission Tracking
±0.018
+0.03/°C
Vista de pagina 78
1 2 ... 74 75 76 77 78 79 80 81 82 83 84 ... 144 145

Comentarios a estos manuales