
3. Preliminary Display Adjustments (SN 3004A and Above)
the combined X deflection voltage applied to the CRT. Use the
oscilloscope front-panel knob to adjust waveform fl sensitivity for
approximately 8 vertical divisions.
hp
running
i
_,..._......_,........................
20.0
V/div
125.000
ns
50.0
ns/div
1
f
25.00 v
Figure 3-27. Composite X Deflection Waveform
17. Adjust
AlA2R217
HF GAIN,
AlA2C204,
and
AlA2C209
for
minimum overshoot and minimum rise and fall times of the
composite
Xdeflection
waveform.
Note
Always adjust
AlA2C204
and
AlA2C209
in approximately equal
amounts. Do not adjust one to its minimum value and the other to its
maximum value.
18. Use the oscilloscope
Cntav)
markers to measure the risetime,
falltime, and percent overshoot of the composite
Xdefection
waveform. Rise and fall times should both be less than
approximately 65 ns between the 10% and 90% points on the
waveform. Overshoot should be less than 3% (approximately 0.25
divisions). See Figure 3-28.
OVERSHOOT
OVERSHOOT
Figure 3-28.
Rise and
I%11
Times and Overshoot Adjustment Waveform
3-60 Adjustments
Comentarios a estos manuales