2 1. Frequency Response Adjustments
70. Set the synthesized sweeper controls as follows:
START FREQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.2.0
GHz
STOP FREQ
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.8
GHz
RF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
LEVELING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .XTAL
SWEEP TIME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.30
ms
SWEEP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CONT
71. On the spectrum analyzer, key in TRACE A [CLEAR-WRITE),
CSWEEP
TIME]
2s,
MARKER (OFF),
IHOLD).
As the spectrum
analyzer completes each sweep, a series of approximately 22 new
responses should be displayed, as shown in Figure 3-95. The
peaks of these responses coarsely outline the spectrum analyzer
frequency response.
Figure 3-95.
Typical Coarse Frequency Response (2
GHz
-
5.8
GHz)
72. Gradually adjust (Band B Lower Segment)
A6AllR51
Bl
and
(Band B Upper Segment)
A6AllR69
B2 for maximum flatness of
the displayed signal responses. The adjustments are interactive,
with
A6AllR51
Bl having the most effect on the level of the
displayed signals below mid-band (approximately 3.9
GHz),
and A6AllR69 B2 having the most effect on the level of
the displayed signals above mid-band. Adjust
A6AllR51
Bl
counterclockwise to increase the level of the displayed signal
responses below mid-band. Adjust A6Al
lR69
B2 clockwise
to increase the level of the displayed signal responses above
mid-band. See Figure 3-90 for the locations of A6AllR51
Bl
and
A6Al
lR69
B2.
Note
It might be helpful to increase or decrease the spectrum analyzer
[SWEEP
TIME) setting while adjusting A6AllR66 B2 and A6AllR48 Bl,
particularly when making fine adjustments. An increase in spectrum
analyzer sweep time results in closer spacing of the displayed
responses, but slows the adjustment.