Agilent Technologies HP 8566B Especificaciones Pagina 207

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 405
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 206
4. Final Display Adjustments (SN 3004A and Above)
Note
Note
9.
10.
11.
AlA2R517 ASTIG
AlA2R513 3D
AlA2R409
FOCUS COMP
Press
(o--2.5),
(REFERENCE LEVEL] and then adjust the reference
level to bring the displayed noise to the top division of the
graticule
((REFERENCE-LEVEL]).
Press
(ENTER~~B/DIV)
and key in 1
dB/DIV. The noise should now completely fill
the
CRT graticule
pattern, illuminating a large rectangular area. If necessary, adjust
the reference level until the graticule pattern is completely filled.
Press SWEEP
CslNGLE),
[SHIFT) DISPLAY LINE
m,
and then
@i!iKj THRESHOLD
(OFFS
to turn off the CRT annotation and
graticule pattern.
Connect a 56503 photometer probe to the Tektronix J-16 digital
photometer. Set the photometer to the Xl range.
Place the photometer light probe hood against the IF-Display
Section glass
RF1
filter, and adjust AlA2R319 INT GAIN for a
photometer reading of 80 NITS
(cd/m2).
This reading must be made with the glass
RF1
filter in place in front of
the CRT. It might be necessary to slightly trim the top and bottom of
the photometer probe’s hood so that it will fit flush against the glass
RF1
filter.
If a standard J-16 photometer is used (instead of metric option
02),
adjust
AlA2R319
for a photometer reading of 23.5 fl (foot-lamberts).
12. Set the spectrum analyzer LINE switch to STANDBY and then
back to ON. The spectrum analyzer power-up annotation should
be visible on the CRT display. (This includes the firmware
datecode.)
13. For the best focus near the center of the CRT display, adjust the
following potentiometers in the sequence listed below. Repeat as
needed to optimize center-screen focus.
AlA3R14
FOCUS LIMIT
AlA2R517 ASTIG
AlA2R513
3D for best focus of annotation (short vectors)
AlA2R409 FOCUS COMP for best focus of graticule lines (long
vectors)
14. Adjust
AlA2R426
T/B
FOC for best focus at the top and bottom of
the display.
15. Adjust
AlA2R437
R/L FOC for best focus at the right and left
sides of the display.
16. If the top and bottom (or right and left sides) of the display
achieve best focus at different potentiometer settings, adjust
AlA2R427 T/B CTR or AlA2R440 R/L CTR, and then readjust
AlA2R426 T/B FOC or AlA2R437
R/L
FOC to optimize overall
focus.
Adjustments 3-67