13. Line-Related Sidebands Test
12. Press MARKER [PEAK SEARCH), [al and position the marker at the
peaks of the line-related sidebands separated from the signal by
multiples of the line frequency; for example, 400 Hz, 800 Hz, 1200
Hz, . . .
13. The MARKER A amplitude for all line sidebands below 2
kHz
should be
c-55
dB.
The A amplitude for sidebands from 2
kHz
to
5.5
kHz
should be
c-65
dB.
Largest Level
<2
kHz
dB
at Hz
Largest Level
<2
kHz
to 5.5
kHz
dB
at
HZ
14. Change (CENTER FREQUENCY) of source analyzer to 5.7
GHz
and
repeat steps 5 through 13 for 5.7
GHz.
Performance Tests 2-57