Power Meter Operation
Making Measurements using Frequency Dependent Offset Tables
Agilent E4418B Users Guide 2-31
you delete the frequency the equivalent offset is also removed and
vice versa.
Note You can only edit the table name you cannot delete it.
6. When you have completed editing the frequency dependent offset
table, press to return to the Offset Tbls screen.
Note Ensure that the frequency points you use cover the frequency
range of the signals you want to measure. If you measure a signal
with a frequency outside the frequency range defined in the
frequency dependent offset table, the power meter uses the
highest or lowest frequency point in the frequency dependent
offset table to calculate the offset.