Agilent Technologies 8487D Guía de usuario Pagina 87

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 196
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 86
Power Meter Operation
Setting Measurement Limits
Agilent E4418B Users Guide 2-47
To set the TTL Output:
1. Press , , . The current settings for the
TTL Output are displayed under the , and
softkeys.
2. To select whether the TTL output reflects an overlimit condition,
under limit condition or both, press . Use any of the four
arrow keys to make a selection between , , and
on the pop up menu that appears.
3. To select whether a high or low level TTL output represents a
limits failure, use the softkey to toggle between
and
TTL Outputs can be disabled and re-enabled simply by pressing
.
Checking for Limit Failures
Limit failures are displayed in the appropriate field in the measurement
window on the power meters display as shown in Figure 2-14.
Note The same limit fail field is used by both channel and window
limits.
Meas
Setup
Limits TTL Output
TTL Output Limits
Fail O/P
Limits
OVER UNDER
EITHER
Fail O/P High
Low
TTL Output Off On
HP4402.book Page 47 Wednesday, November 29, 2000 4:46 PM
Vista de pagina 86
1 2 ... 82 83 84 85 86 87 88 89 90 91 92 ... 195 196

Comentarios a estos manuales

Sin comentarios