Agilent Technologies 8487D Guía de usuario Pagina 81

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 196
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 80
Power Meter Operation
Measuring Pulsed Signals
Agilent E4418B Users Guide 2-41
Measuring Pulsed Signals
The power meter can be used to measure the power of a pulsed signal. The
measurement result is a mathematical representation of the pulse power
rather than an actual measurement (assumes constant peak power). The
power meter measures the average power of the pulsed input signal and
then divides the measurement result by the duty cycle value to obtain the
pulse power reading. The allowable range of values is 0.001 % to 100 %.
The default value is 1.000%.
If duty cycle is enabled then Dty Cyc is displayed.
Note Pulse measurements are not recommended using Agilent E4412A
and E4413A power sensors.
An example of a pulsed signal is shown in Figure 2-10.
Figure 2-10: Pulsed Signal
Power
Time
B
A
Duty Cycle = A
B
HP4402.book Page 41 Wednesday, November 29, 2000 4:46 PM
Vista de pagina 80
1 2 ... 76 77 78 79 80 81 82 83 84 85 86 ... 195 196

Comentarios a estos manuales

Sin comentarios