Agilent Technologies 8935 Series Especificaciones Pagina 179

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 343
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 178
178 Chapter 8
Performance Tests
RF Generator FM Distortion Performance Test 1
RF Generator FM Distortion
Performance Test 1
The purpose of this test is to verify that the Test Set meets the
specification limits in PTR (Performance Test Record) Table 9-1, “RF
Generator FM Distortion Test 1 Record,” on page 234. The FM
distortion of the RF generator is measured directly by the measuring
receiver. The Test Set’s internal audio generator provides the
modulation source.
NOTE
Two setups are shown below. The first setup can measure signals to 1
GHz. Since the FM generator in the Test Set translates FM in the lower
band directly into the 1.7 to 2 GHz range, testing to 1 GHz is adequate
when verifying a repair. The second setup has a microwave converter
which covers the full measurement range of FM signals to 2 GHz.
Initial Setup
Figure 8-1 Setup for Measurements to 1 GHz
Figure 8-2 Setup for Measurements to 2 GHz Using a Microwave Converter
INPUT 50
DUPLEX OUT
Measuring Receiver
INPUT 50
DUPLEX OUT
Measuring Receiver
Microwave Converter
Signal Generator
IF OUTPUT
RF INPUT
LO INPUT
RF OUTPUT
Vista de pagina 178
1 2 ... 174 175 176 177 178 179 180 181 182 183 184 ... 342 343

Comentarios a estos manuales

Sin comentarios