138 Chapter 6
O:\Manuals\E6380A_Progguid\Book\Advoper.fm
Advanced Operations
Status Reporting
Table 6-3 Standard Event Status Register Bit Assignments
Bit
Number
Binary
Weighting
Condition Comment
15 32879 Always 0 Reserved by IEEE 488.2
14 16384 Always 0 Reserved by IEEE 488.2
13 8192 Always 0 Reserved by IEEE 488.2
12 4096 Always 0 Reserved by IEEE 488.2
11 2048 Always 0 Reserved by IEEE 488.2
10 1024 Always 0 Reserved by IEEE 488.2
9 512 Always 0 Reserved by IEEE 488.2
8 256 Always 0 Reserved by IEEE 488.2
7128Power On1 = Test Set’s power supply has been turned off and
then on since the last time this register was read.
664User
Request
Not implemented in Test Set.
5 32 Command
Error
1 = The Test Set detected an error while trying to
process a command. The following events cause a
command error:
An IEEE 488.2 syntax error. This means that
the Test Set received a message that did not
follow the syntax defined by the Standard.
A semantic error. For example, the Test Set
received an incorrectly spelled command.
The Test Set received a Group Execute Trigger
(GET) inside a program message.
416Execution
Error
1 = The Test Set detected an error while trying to
execute a command. The following events cause an
execution error:
A <PROGRAM DATA> element received in a
command is outside the legal range for the Test
Set or is inconsistent with the operation of the
Test Set.
The Test Set could not execute a valid command
due to some Test Set hardware/firmware
condition.