Agilent Technologies 8935 Series Manual de usuario Pagina 160

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 329
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 159
160 Chapter 6
O:\Manuals\E6380A_Progguid\Book\Advoper.fm
Advanced Operations
Status Reporting
Table 6-8 shows the Hardware Status Register Group #1’s Condition
Register bit assignments.
Table 6-8 Hardware Status Register Group #1, Condition Register Bit
Assignments
Bit
Number
Binary
Weighting
Condition Comment
15 32768 Not Used (Always 0) Defined by SCPI Version 1994.0
14 16384 Radio Interface Card
Interrupt #2 Tripped
13 8192 Radio Interface Card
Interrupt #1 Tripped
12 4096 Signaling Decoder
Measurement Results
Available
11 2048 Signaling Decoder Input
Level Too Low
10 1024 Signaling Decoder is
Measuring
9 512 Signaling Decoder is Armed
8 256 Signaling Encoder Sending
Auxiliary Information
If the Signaling Mode selected has
two information fields, such as the
AMPS Filler and Message fields,
and both fields are being sent, this
bit will be set.
7 128 Signaling Encoder Sending
Information
If the Signaling Mode selected has
only one information field and the
field is being sent, this bit will be set
high. If the Signaling Mode selected
has two information fields, such as
the AMPS Filler and Message
fields, and only one field is being
sent, this bit will be set high. This
bit is not active if the Signaling
Encoder Mode is set to Function
Generator.
6 64 Communication Register
Group Summary Message
1 = one or more of the enabled
events have occurred since the last
reading or clearing of the Event
Register.
5 32 Measurement Limit(s)
Exceeded
This bit is set high if the
Measurement High Limit or Low
Limit is exceeded.
4 16 Power-up Self Test(s) Failed
Vista de pagina 159
1 2 ... 155 156 157 158 159 160 161 162 163 164 165 ... 328 329

Comentarios a estos manuales

Sin comentarios