144 Chapter 6
O:\Manuals\E6380A_Progguid\Book\Advoper.fm
Advanced Operations
Status Reporting
Accessing the Output Queue
When messages are sent to the Test Set, it decodes the message to
determine what commands have been sent. Depending upon the type of
command, the Test Set’s processor sends messages to various parts of
the instrument. Many commands generate data which must be sent
back to the controller. This data is buffered in the Output Queue until it
is read by the controller. The availability of data is summarized in the
MAV bit of the Status Byte Register. The state of the MAV message
indicates whether or not the Output Queue is empty. The MAV message
is TRUE, logic 1, when there is data in the Output Queue and FALSE,
logic 0, when it is empty. The Output Queue is read by addressing the
Test Set to TALK and then handshaking the bytes out of the Output
Queue. Depending upon the type of command sent, the data may
appear in the Output Queue almost immediately, or it may take several
seconds (as is the case with some Signaling Decoder measurements).
Care should be exercised when reading the Output Queue since the
HP-IB bus will, by design, wait until the data is available before
processing further bus messages.
Reading the Output Queue
Example 6-40 Example
Enter 714;Output_data