Test Re cords Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 153
Agilent 81600B #200 Performance Test Page 6 of 14
Model Agilent 81600B #200 Tunable Laser Report No. ________ Date_______
Wavelength Repeatability
Measurement Uncertainty: ± 0.1 pm
Repeatability of
1440.000 nm (= reference)
Measurement Result
Repeatability of
1540.000 nm (= reference)
Measurement Result
Initial Setting REF = nm Initial Setting REF = nm
from 1490.000 nm to REF nm from 1440.000 nm to REF nm
from 1540.000 nm to REF nm from 1490.000 nm to REF nm
from 1590.000 nm to REF nm from 1590.000 nm to REF nm
from 1640.000 nm to REF nm from 1640.000 nm to REF nm
largest measured wavelength nm largest measured wavelength nm
smallest measured
wavelength
nm smallest measured
wavelength
nm
Wavelength Repeatability nm Wavelength Repeatability nm
= largest measured wavelength - smallest measured
wavelength
= largest measured wavelength - smallest measured
wavelength
Upper Test Limit 0.0016 nm Upper Test Limit 0.0016 nm
Performance Characteristic 0.0010 nm typical Performance Characteristic 0.0010 nm typical
Repeatability of
1640.000 nm (= reference)
Measurement
Result
Initial Setting REF =
nm
from 1440.000 nm to REF nm
from 1490.000 nm to REF nm
from 1540.000 nm to REF nm
from 1590.000 nm to REF nm
largest measured wavelength nm
smallest measured
wavelength
nm
Wavelength Repeatability nm
= largest measured wavelength - smallest measured
wavelength
Upper Test Limit 0.0016 nm
Performance Characteristic 0.0010 nm typical