Test Re cords Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 209
Agilent 81600B #130 Performance Test Page 5 of 12
Model Agilent 81600B #130 Tunable Laser Report No. ________ Date_______
Mode Hop Free Tuning
1
Wavelength Deviation = Wavelength Measured - Wavelength Setting
Mode Hop Free Tuning Result (= Largest value of either the Maximum or Minimum Deviation)
Mode Hop Free Tuning Result: _______ nm
Te s t L i mi t : ±0.025 nm
Measurement Uncertainty: ±0.2 pm
Wavelength Setting Wavelength Measured Wavelength Deviation
1
1260.000 nm nm nm
1261.000 nm nm nm
1262.000 nm nm nm
1263.000 nm nm nm
1264.000 nm nm nm
1265.000 nm nm nm
1266.000 nm nm nm
1267.000 nm nm nm
1268.000 nm nm nm
1269.000 nm nm nm
1270.000 nm nm nm
1365.000 nm nm nm
1366.000 nm nm nm
1367.000 nm nm nm
1368.000 nm nm nm
1369.000 nm nm nm
1370.000 nm nm nm
1371.000 nm nm nm
1372.000 nm nm nm
1373.000 nm nm nm
1374.000 nm nm nm
1375.000 nm nm nm
Maximum Deviation: nm
Minimum Deviation: nm