Making Measurements 2
Measuring Capacitance
U1231A/U1232A/U1233A User’s Guide 49
Measuring Capacitance
Set up your multimeter to measure capacitance as shown in
Figure 2- 17. Probe the test points and read the display.
Figure 2-16 Capacitance display
CAUTION
To avoid possible damage to the multimeter or to the equipment under
test, disconnect circuit power and discharge all high-voltage
capacitors before measuring capacitance. Use the DC voltage function
to confirm that the capacitor is fully discharged.
NOTE
The multimeter measures capacitance by charging the capacitor with a
known current for a known period of time, measuring the resulting
voltage, and then calculating the capacitance.
is shown on the left of the display when the capacitor is charging,
and is shown when the capacitor is discharging.
NOTE
To improve measurement accuracy of small value capacitors, press
with the test leads open to subtract the residual capacitance of
the multimeter and leads.
For measuring capacitance values greater than 1000 μF, discharge the
capacitor first, then select a suitable range for measurement. This will
speed up the measurement time and also ensures that the correct
capacitance value is obtained.