2 Making Measurements
Measuring Temperature
56 U1231A/U1232A/U1233A User’s Guide
Available options:
°C - Temperature measured in °C only.
°C°F - During temperature measurements, press to
switch between °C and °F.
°F - Temperature measured in °F only.
°F°C - During temperature measurements, press to
switch between °F and °C.
4 Press to save the changes. Press and hold until
the multimeter restarts
Temperature measurement without ambient compensation
If you are working in a constantly varying environment,
where ambient temperatures are not constant, do the
following:
1 Press and hold for more than 1 second to select the
0 °C compensation ( ). This allows a quick measurement
of the relative temperature.
2 Avoid contact between the thermocouple probe and the
surface to be measured.
3 After a constant reading is obtained, press to set the
reading as the relative reference temperature.
4 Touch the surface to be measured with the thermocouple
probe and read the display.
Figure 2-21 Temperature measurement without ambient compensation
A
u
to
R
a
n
g
e
A
u
to
R
a
n
g
e
A
u
t
o
T
ri
H
o
l
d
g
L
o
g
S
h
i
f
t
E
s
c
A
u
t
o
R
a
n
g
e
N
u
l
l
R
e
c
a
l
l
Vista de pagina 73
1 2 ... 69 70 71 72 73 74 75 76 77 78 79 ... 132 133

Comentarios a estos manuales

Sin comentarios