Clearing Non-Volatile Memory
NOTE It is important to Clear the HP/Agilent 8922 non-volatile RAM before
and after the firmware upgrade.
To clear non-
volatile RAM
Step 1. Move the cursor to the CONFIG field and select.
Step 2. In the CONFIGURE screen, move the cursor to the SERVICE field and
select.
Step 3. In the SERVICE screen, locate the RAM Initialize field and select.
Step 4. Select on the front panel keypad to clear the registers.
YES