Calibrating the Internal Reference
To complete the ROM upgrade, you must recalibrate the internal
reference.
NOTE Before recalibrating the internal reference, allow the instrument to
warm up for at least 10 minutes. It must be performed after replacement
of the reference section A15, or if the instrument displays the error
message
Frequency reference cal lost. Perform reference calibration.
Step 1. Connect a reference 10 MHz source to the rear panel REF IN.
Step 2. On the Configuration screen, select the field and highlight Calibrate.
Step 3. HP/Agilent 8922A/B only: At power up, select . This
takes you to the Configure screen. At the Configure screen, select
Reference and highlight Calibrate.
HP/Agilent 8922E/F/G/H/S only: At turn on, the instrument is at the
Cell Configuration or Cell Status screen. Move the cursor to the To
Screen section and select More. This takes you to the Choices section.
Select CONFIG. At the Configure screen, select Reference and highlight
Calibrate.
After approximately 15 seconds the reference is calibrated.
Verifying correct upgrade
After the frequency reference is calibrated, cycle the instrument power. If
successfully calibrated, the instrument powers up with the message All
Self Tests Passed only.