lpErrorDoc.txt
fragmentation). Report this error to Agilent Technologies. The
instrument is still functional.
-321 Out of memory; Unable to display timeslot window.
The instrument was unable to create part of the graphical user
interface due to an inability to allocate memory (possibly due to
fragmentation). Report this error to Agilent Technologies. The
instrument is still functional.
-321 Out of memory; User Flatness subsystem out of memory.
There is insufficient working memory available. Report this error
to Agilent Technologies.
-321 Out of memory; User Power Correction subsystem out of memory.
There is insufficient working memory available. Report this error
to Agilent Technologies.
-321 Out of memory; {name}
The header file to be written is too big to fit on the media. The
new information was not written and the original header is
untouched.
-321 Out of memory; {name} display cannot function.
There is insufficient working memory available. Report this error
to Agilent Technologies.
-315 Configuration memory lost; Persistent state checksum is bad. Using
factory defaults.
The persistent state is corrupt and had to be reinitialized with
the factory preset values.
-315 Configuration memory lost; Persistent state preset. Using factory
defaults.
The persistent state has been forced to return to factory preset
values.
-313 Calibration memory lost; Cal not found and was created: {name}
The specified cal file was not found so it was created by the
system.
-313 Calibration memory lost; Cal out of mem: {name}
Inadequate memory for specified cal file.
-313 Calibration memory lost; Cal resizing: {name}
Inadequate memory to resize specified cal file.
-313 Calibration memory lost; Cal restored from backup: {name}
A specified cal file was restored from backup.
-313 Calibration memory lost; Cannot restore cal from backup:
{name}err={value}
Page 5