E-TFCI value should be 0 to 125 when E-TFCI table is 1 and TTI is
2msec.
E-TFCI value shoudl be 0 to 120 when E-TFCI table is 1 and TTI is
10msec.
-221 Settings conflict; EDGE BERT requires multiframe PN Sequence in
Timeslot E field.
An unsupported data type was encountered. The measurement timeslot
requires a multiframe PN sequence in the E field.
-221 Settings conflict; EDGE BERT requires Normal as Timeslot Type.
An unsupported timeslot type was encountered. The measurement
timeslot must be designated Normal.
-221 Settings conflict; Enabled modulation source conflicts with previously
enabled modulation source. Previous modulation disabled.
Three sources are shared by FM1/PM1, AM1/AM2, FM2/PM2, pulse and
burst envelope: INT, EXT1, and EXT2. Each source can be used by
only one modulation at a time. If a modulation requests a source
while it is in use by an active modulation, the first modulation is
turned off, and the second modulation is turned on.
-221 Settings conflict; External Ack/Nack delay(EADelay) value is too
large.
External Ack/Nack delay(EADelay) value should be 0 to 61184 when
TTI is 2msec.
-221 Settings conflict; External data cannot be selected for multiple
channels. PN9 is automatically selected for {name}.
The current hardware only supports one external data source at any
given time. The <CHANNEL NAME> data source has been changed to PN9,
the default data source.
-221 Settings conflict; FFCH/OCNS and QPCH cannot be on at the same time.
QPCH has been turned off.
QPCH is generated using both channel 3 and 4. Channel 3 has been
turned on to generate FFCH or OCNS. Channel 4 can no longer
generate QPCH. Therefore, QPCH must be turned off.
-221 Settings conflict; FM and PM not allowed.
There is a hardware conflict between FM and PM. The most recently
requested modulation will be turned on. The previous modulation
will be turned off.
-221 Settings conflict; FM1/PM1 value set less than FM2/PM2 value. FM2/PM2
changed to match FM1/PM1 value.
The deviation of FM2/PM2 must always be less than or equal to the
deviation settings for FM1/PM1. This error will be reported to the
queue when FM2/PM2 is enabled and FM1/PM1 is also enabled and an
adjustment to either FM1/PM1 deviation causes the FM1 or PM1
deviation to be less than the FM2 or PM2 deviation. It will also be
reported when FM1/PM1 is being turned on, and the last FM2/PM2
deviation setting is greater than the current FM1/PM1 deviation
setting. In both cases the FM2/PM2 deviation will be adjusted to