Agilent Technologies EXG Especificaciones Pagina 237

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 358
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 236
$JLOHQW0$996HULHV6HUYLFH*XLGH
5HSDLULQJWKH$JLOHQW0RGHO9&79&7
Repairing the Agilent
Model V24CT/V26CT
Rear Housing Removal Procedures
Disassembling the Rear Housing consists of:
Removing power to the Model V24CT/V26CT by pressing the power switch to OFF and
disconnecting the AC power cord.
Removing Lead Acid Batteries (see 9-5).
Removing Parameter Module Rack (see 9-6).
Removing the Front Housing Assembly (see 9-2).
Removing Battery Contact Board (see 9-7).
Removing the System Board (see 9-8).
Removing Power Supply Assembly (see 9-10).
Removing Parameter Module Rack Connector (see 9-13).
5HPRYLQJ%DWWHU\&RQWDFW%RDUG
Remove power to the Model V24CT/V26CT by pressing the power switch to OFF and
disconnecting the AC power cord.
Remove batteries. See 9-5.
Remove Front Housing Assembly. See 9-2.
Remove battery contact board power cable connector.
Remove four screws securing battery contact board to power supply assembly
Remove battery contact board.
Contact Board
Power Cable
Connector
Vista de pagina 236
1 2 ... 232 233 234 235 236 237 238 239 240 241 242 ... 357 358

Comentarios a estos manuales

Sin comentarios