$JLOHQW0$996HULHV6HUYLFH*XLGH
)URQW+RXVLQJ$VVHPEO\9&9&9&79&7
Front Housing Assembly
V24C/V26C & V24CT/
V26CT
5HSODFLQJWKH%RDUG6KLHOG
Place the Board Shield on the Flat Panel Bracket.
Screw in the 2 hex stand-off screws (5mm) located in the narrow channel securing the
Board Shield and Flat Panel Bracket to the top of the Front Housing.
Screw on the 4 hex nuts (5.5mm) securing the rest of the Board Shield to the Flat Panel
Bracket.
5HSODFLQJ3DUDPHWHU0RGXOH5DFN$VVHPEO\0
Align the slides on the Front End Rack Assembly with the mating slides on the monitor.
Slide the Front End Rack Assembly downward until it snaps into place.
Parameter modules can now be inserted into the parameter module rack assembly by slid-
ing them into the rack until they snap into place.
5HSODFLQJ/HDG$FLG%DWWHULHV&7PRGHO
Remove power to the instrument by pressing the power switch to OFF and disconnecting
the AC power cord.
At rear of the instrument, open battery door (swings downward).
Slide lead-acid battery into chamber until it snaps into place. Repeat for other battery if
necessary.