$JLOHQW0$996HULHV6HUYLFH*XLGH
)URQW+RXVLQJ$VVHPEO\9&9&9&79&7
Front Housing Assembly
V24C/V26C & V24CT/
V26CT
Lay the flex cable assembly over the edge of the Front Housing.
Disconnect the 12V cable connecting the CDCI Board to the Flat Panel Adapter Board.
Remove the 3-Board assembly by:
a. Removing the 3 screws securing the middle DPSC board to the Flat Panel Bracket.
b. Removing the six screws securing the CDCI Board and the Flat Panel Adapter Board
to the Flat Panel Bracket.
c. Lifting the entire 3-board assembly off the Flat Panel assembly.
If necessary to remove individual boards from the board assembly, grip either the CDCI
Board or the Flat Panel Adapter Board securely and gently pull so that the board connec-
tor separates from the DSPC board.