Agilent Technologies Agilent G1369B Manual de usuario Pagina 128

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 206
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 127
128 Agilent 1260 Infinity High Performance Autosampler User Manual
9 Maintenance
Installing the needle assembly
1
Push the silicon safety tube delivered with every needle
over the needle.
2 Insert the loop capillary into the needle assembly and
tighten the fitting hand tight.
3 Pinch the holder clamp and reinsert the needle assembly
into the needle carrier.
4 Attach a 5/16 inch wrench to hold the position at the
needle assembly. Use a 1/4 inch wrench to tighten the
fitting of the loop capillary.
Vista de pagina 127
1 2 ... 123 124 125 126 127 128 129 130 131 132 133 ... 205 206

Comentarios a estos manuales

Sin comentarios