Agilent 1260 Infinity High Performance Autosampler User Manual 129
Maintenance
9
Installing the needle assembly
5
Close the leak guide
Next Steps:
6 Check the alignment of the needle in the needle pusher
of the needle carrier by viewing from several directions to
see that it is aligned in the center of the needle pusher.
NOTE
The needle must be centered in the needle pusher as
all alignment by the Autosampler is calculated from
the needle pusher position.
7
Remove the silicon safety tube from the needle.
8 In the user interface exit the Change needle/seat
function and exit the maintenance mode. In the Lab
Advisor software the Change needle/seat function can
be found in the Tools section.
9 Re-install the side door and close the front door.