Agilent Technologies E1345A Manual de usuario Pagina 50

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 93
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 49
Relay Multiplexer Command Reference 51
ARM
The ARM subsystem selects the number of scanning cycles (1 through 32767) for
each
INITiate command.
ARM
:COUNt <
number> MIN|MAX
:COUNt? [MIN |MAX]
:COUNt
ARM:COUNt <number> MIN|MAX allows scanning cycles to occur a multiple
of times (1 to 32767) with one
INITiate command and when
INITiate:CONTinuous OFF|0 is set.
Parameters
Comments Number of Scans: Use only values between 1 to 32767 for the number of
scanning cycles (default is 1).
Related Commands: ABORt, INITiate[:IMMediate]
*RST Condition: ARM:COUNt 1
Example Setting Ten Scanning Cycles
ARM:COUN 10
Sets 10 scanning cycles.
SCAN (@100:115)
Sets channel list.
IINIT
Starts scanning cycle.
Parameter Name Parameter Type Range of Values
<
number>
numeric 1–32767|MIN|MAX
Vista de pagina 49
1 2 ... 45 46 47 48 49 50 51 52 53 54 55 ... 92 93

Comentarios a estos manuales

Sin comentarios