Agilent Technologies B2201A Manual de servicio Pagina 103

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 124
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 102
Module 13
SPGU Control and Applications
13-17
p
n+
n+
Source
Drain
Gate
Substrate
Icp
SMU
SMU
PGU
or
Oxide
A
A
Charge Pumping
Charge pumping is a type of hot carrier measurement. It provides direct measurement of interface
states and an indication of electron and hole trapping.
The gate of the MOS transistor is connected to a pulse generator. The current (Icp) is caused by the
repetitive recombination of minority carriers with majority carriers at the silicon-silicon oxide
interface.
Vista de pagina 102
1 2 ... 98 99 100 101 102 103 104 105 106 107 108 ... 123 124

Comentarios a estos manuales