Module 11
Advanced Definitions and Operations
11-16
Final_xxxx.exe
Start
End
Chuck down
Arguments: -a GPIB_address
-l log_file_name
Example:
–a GPIB0::5::INSTR –l C:\temp\prb.log
Response: none
C:\Program Files\Agilent\B1500\EasyEXP
C:\Program Files\Agilent\B1500\EasyEXP
C:\Program Files\Agilent\B1500\EasyEXP
5
Stop condition
To specify the final procedure on the Repeat Measurement Setup dialog box, enter the full path name
of Final_xxxx.exe into the Final Procedure field.
The procedure is called after one of the stop condition is detected. This procedure sets wafer chuck
to the DOWN position.
Note: To ignore the Arguments, set the GPIB address and log file name in the prober_info.ini.