Language Dictionary - 3
57
OUTPut:IMPedance
Agilent 6811B, 6812B, 6813B, Only
This command enables or disables the output impedance programming capability of the ac source.
Command Syntax
OUTPut:IMPedance[:STATe] <bool>
Parameters
0 | 1 | OFF | ON
*RST Value
OFF
Examples
OUTP:IMP 1 OUTP:IMP ON
Query Syntax
OUTPut:IMPedance[:STATe]?
Returned Parameters
0 | 1
Related Commands
OUTP:IMP:REAL OUTP:IMP:REAC
OUTPut:IMPedance:REAL
Agilent 6811B, 6812B, 6813B, Only
This command sets the real part of the output impedance of the ac source. OUTPut:IMPedance:STATe
must be enabled for the programmed impedance to affect the output.
Command Syntax
OUTPut:IMPedance:REAL <NRf>
Parameters
0 to 1 (ohms)
*RST Value
0
Examples
OUTP:IMP:REAL 0.25
Query Syntax
OUTPut:IMPedance:REAL?
Returned Parameters
<NR3>
Related Commands
OUTP:IMP OUTP:IMP:REAC
OUTPut:IMPedance:REACtive
Agilent 6811B, 6812B, 6813B, Only
This command sets the reactive part of the output impedance of the ac source.
OUTPut:IMPedance:STATe must be enabled for the programmed impedance to affect the output.
Command Syntax
OUTPut:IMPedance:REACtive <NRf>
Parameters
0.00002 to 0.001 (henrys)
*RST Value
0.0005
Examples
OUTP:IMP:REAC 100E-6
Query Syntax
OUTPut:IMPedance:REAC?
Returned Parameters
<NR3>
Related Commands
OUTP:IMP OUTP:IMP:REAL