Chapter 3: Testing Performance
To test the threshold accuracy
Connect and configure the logic analyzer
1 Using the 17-by-2 test connector, BNC cable, and probe tip assembly,
connect the data and clock channels of Pod 1 to the free side of the BNC
Tee.
2 Configure the logic analyzer:
a Click the Bus/Signal Setup icon. The Analyzer Setup dialog opens.
b In the Buses/Signals tab, click Delete All at the bottom of the dialog.
c Using the mouse, activate all Pod 1 channels. Assign channels to bus/
signal name My Bus 1.
d Scroll the channel assignments to the left. Assign the clock/data
channel for the Pod 1 (that is, C1) to My Bus 1.
3 Activate the DC source output.
Vista de pagina 28