Chapter 5: Troubleshooting
Troubleshooting the Agilent 1680A,AD-series
Possible problem with
logic analyzer cables. Do
the cable test in Chapter
5 on suspect pod.
Does cable
test pass?
Swap suspect probe
tip assembly with
known good one.
Does cable
test pass?
Swap suspect cable
assembly with known
good one.
Does cable
test pass?
Replace logic analyzer
acquisition board.
Replace
defective cable.
Replace defective
probe tip assembly.
The logic analyzer
board if functioning
properly.
No
Yes
No
Yes
No
Yes
7
1