Chapter 3: Testing Performance
To test the single-clock, single-edge, state acquisition
Connect the 1682/83/92/93A,AD Logic Analyzer to the Pulse Generator
3 Activate the data channels that are connected according to one of the
previous tables:
a Click on the Bus/Signal Setup icon. The Analyzer Setup dialog opens.
b In the Buses/Signals tab, click Delete All at the bottom of the dialog.
c Using the mouse, activate the data channels being tested. Assign
channels to bus/signal name My Bus 1.
d Click OK to close the Analyzer Setup dialog.
Test i n g
Combination
Connect to 8133A
Channel 2 Output
Connect to 8133A
Channel 2 Output
Connect to 8133A
Channel 1 Output
1 Pod 1, channel 3
Pod 2, channel 3
Pod 3, channel 3
Pod 4, channel 3
Pod 1, channel 11
Pod 2, channel 11
Pod 3, channel 11*
Pod 4, channel 11*
Pod 1 clock/data channel (Clk 1)
*1682A,AD or 1692A,AD only.