Increment This softkey is displayed only when Save To Trace is selected.
When On, all test violations are saved by incrementing the trace
number. The starting trace number is the one that is currently
selected. When the 64K compressed memory is full, the oldest
trace memory is overwritten, and the trace count continues
incrementing. When the trace count reaches 100, the number
resets to 1 (wraps around).
When Off, only the last test violation is saved, as test failure data is
written over previously stored data.
Previous Menu Returns you to the previous softkey menu.
5
Toggle the softkeys to select the desired testing options.
6 Press the Previous Menu softkey.
7 Press the Mask Test softkey until On is selected.
Selecting on immediately starts the test using the test options specified. Test
indications are displayed on the display line as follows.
Pass Indicates the displayed waveform passed the test.
Fail Indicates the displayed waveform failed the test. Further testing,
and disposition of the failed data is dependent on the testing options
selected.
Acquisitions Indicates the total acquisitions made during the test.
Failures Indicates the total number (and percentage) of test failures that
occurred during the test.
Waveform Monitoring Operating Hint
Mask template testing can only be used in the Main Horizontal Mode, and when
Functions 1 and 2 are set to off.
The trace review softkey can be used to review all saved failures. See "To Save
or Recall Traces" in this chapter for more information.
Operating the Measurement/Storage Module
To start waveform monitoring
2–39