To automatically save test violations
The signals that fail the waveform monitoring test can be saved, then
viewed/measured at a later time. Provisions are provided to save the
violations in trace memory, or print a hardcopy of the data. When trace is
selected, the option of saving only the last violation, or saving all violations
are provided.
1
Setup for waveform monitoring as described previously.
2 Press
± .
3 Press the Mask Test softkey.
4 Press the Test Options softkey.
5 Press the Auto Save softkey to ON. This causes the Save To softkey to appear.
Determine how the test violations are being saved, and proceed as follows:
To print test violation data on an externally connected printer, toggle the
Save To softkey until Print is selected.
To save test violation data in trace memory, toggle the Save To softkey
until Trace is selected. This causes the Increment softkey to appear.
To save only the last test violation waveform in trace memory, toggle the
Increment softkey until Off is selected.
To save all test violation waveforms in trace memory, toggle the Increment
softkey until On is selected.
6
Press the Previous Menu softkey.
7 Press the Mask Test softkey until On is selected.
Operating the Measurement/Storage Module
To automatically save test violations
2–40