Chapter 5: Replacing Assemblies
To remove and replace the on/off board
101
Figure 5-18
This figure shows the other left hand side tab for the control panel faceplate. This view is with the oscilloscope turned on its
side and looking at this tab from underneath the instrument.
8
To replace the front panel assembly, reverse the above procedure. Suggestion: When
you get to the part where you have to snap the tabs located on the control panel
faceplate back into the chassis, it is best to snap the faceplate and the assembly together
first before trying to attach them back to the chassis via these tabs. Once they are
attached together, snap down the right hand side tabs first and then the left hand tabs
second.
To remove and replace the on/off board
Use this procedure to remove and replace the on/off board. When necessary, refer to other
removal procedures.
1
Remove the front panel bezel (directions in the previous section).
2
Remove the one Torx T10 screw from the on/off board (Figure 5-19).
Figure 5-19
3
Pull the board out a little so you can see the cables connected behind it (Figure 5-20).
Disconnect these cables and then take the board off.