Chapter 5: Replacing Assemblies
To remove and replace the power cord
124
To remove and replace the power cord
Use this procedure to remove and replace the power cord. When necessary, refer to other
removal procedures. The graphics in this chapter are representative of the oscilloscope at the
time of this printing. Your unit may look different.
1
Disconnect the power cable and remove the cover and top plate.
2
Unplug the power cable from the 12V power supply (refer to Figure 5-52 for the location
of this connection).
3
Unscrew the power cord connector from rear panel (Torx T10 screws) - see Figure 5-54.
Figure 5-54
4
Remove the ground screw (Torx T10) on right side of hard drive (Figure 5-55).
Figure 5-55
5
Pull the power cord out of the instrument.
6
To replace the power cord, reverse these directions.