Chapter 3: Testing Performance
Performing Self-Test and Calibration
19
Performing Self-Test and Calibration
1
Perform self tests
a Pull down the Utilities menu and select Self Test.
b Select Scope Self Test from the Self Test list.
c Click on Start Self Test to start the self test procedure.
If any of the self-tests fail, ensure that the failure is diagnosed and repaired before
calibrating and testing performance.
2
Perform calibration. See “To run the self calibration” on page 15.