122 Programmer’s Guide
5 Command Reference
SEM Offset Sweep Time
[:SENSe]:SEMask:OFFSet:LIST:SWEeptime
<time>,<time>,<time>,<time>,<time>
[:SENSe]:SEMask:OFFSet:LIST:SWEeptime?
Use this command to specify the related sweep time of offset channels
in SEM measurement. The unit for the time in this command is second.
[:SENSe]:SEMask:OFFSet:LIST:SWEeptime:AUTO
OFF|ON|0|1,...
[:SENSe]:SEMask:OFFSet:LIST:SWEeptime:AUTO?
Use this command to turn on/off automatically setting of sweep time of
reference channels in SEM measurement.
Example: :SENSe:SEMask:OFFSet:LIST:SWEeptime
10,10,10,10,10
Key access: Meas > SEM > Offset/Limit > Sweep Time
SEM Offset Abs StartAmpt
[:SENSe]:SEMask:OFFSet:LIST:STARt:ABSolute
<ampl>,<ampl>,<ampl>,<ampl>,<ampl>
[:SENSe]:SEMask:OFFSet:LIST:STARt:ABSolute?
Use this command to specify the start frequency amplitude of offset
channel absolute limit mask in SEM measurement. Before use this
command, the fail mask type should be set to “Absolute” or “Abs AND
Rel” or “Abs OR Rel”.
Example: SENSe:SEMask:OFFSet:LIST:STARt:ABSolute -14,
-14,-15,-20,-20
Key access: Meas > SEM > Offset/Limit > More 1 of 3 > Abs StratAmpt