Command Reference 5
Programmer’s Guide 81
Bandwidth Subsection
Resolution Bandwidth
[:SENSe]:BANDwidth|BWIDth[:RESolution] <freq>
[:SENSe]:BANDwidth|BWIDth[:RESolution]?
Specifies the resolution bandwidth. For numeric entries, all RBW types
choose the nearest (arithmetically, on a linear scale, rounding up)
available RBW to the value entered.
Range: 10 Hz to 3 MHz
Default Unit: Hz
Example: :BAND 1 kHz
Key access: BW > RBW
Auto Resolution Bandwidth State
[:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO
OFF|ON|0|1
[:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO?
This command turns on/off auto resolution bandwidth state.
*RST: On
Example: :BWID:AUTO On
Video Bandwidth
[:SENSe]:BANDwidth|BWIDth:VIDeo <freq>
[:SENSe]:BANDwidth|BWIDth:VIDeo?
Specifies the video bandwidth.
*RST: 3 MHz
Range: 1 Hz to 3 MHz. This range is dependent upon the setting of
[:SENSe]:BANDwidth|BWIDth[:RESolution].
Default Unit: Hz
Example: :BANDwidth:VIDeo 10 KHZ
Key access: BW > VBW Auto Man