Agilent Technologies N9344C Manual de servicio Pagina 141

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 224
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 140
Command Reference 5
Programmer’s Guide 133
Spectrogram Save File Type
[:SENSe]:SPECtrogram:SAVE:FTYPe TRC|CSV
[:SENSe]:SPECtrogram:SAVE:FTYPe?
Use this command to switch the spectrogram test result type between
TRC and CSV.
Example: :SPECtrogram:SAVE:FTYPe TRC
Key access: Meas > File Logging > File Type
Spectrogram Save File Size
[:SENSe]:SPECtrogram:SAVE:FSIZe <int>
[:SENSe]:SPECtrogram:SAVE:FSIZe?
Use this command to set the maximum number of trace (frame) saved in
one spectrogram test result file.
Range: 1 to 1500 frames
Example: :SPECtrogram:SAVE:FSIZe 1400
Key access: Meas > Spectrum Monitor > File Logging > File Size
Spectrogram Marker Y Value
[:SENSe]:SPECtrogram:MARKer[1]|2:Y?
Use this command to query the Marker Y axis measurement result in
spectrogram measurement.
Example: :SPECtrogram:MARKer:Y?
Vista de pagina 140