Agilent Technologies N9344C Manual de servicio Pagina 138

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 224
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 137
130 Programmer’s Guide
5 Command Reference
Spectrogram Marker Band Power State
[:SENSe]:SPECtrogram:MARKer:BPOWer:STATe
OFF|ON|0|1
[:SENSe]:SPECtrogram:MARKer:BPOWer:STATe?
Use this command to turn on/off the band power test of the two Markers
in spectrogram measurement.
Example: :SPECtrogram:MARKer:BPOWer:STATe ON
Key access: Marker > More 1 of 2 > Band Power
Spectrogram Marker Band Power Value
[:SENSe]:SPECtrogram:MARKer:BPOWer?
Use this command to query the Marker band power measurement result.
Example: :SPECtrogram:MARKer:BPOWer?
Spectrogram Marker Band Width Value
[:SENSe]:SPECtrogram:MARKer:BWIDth?
Use this command to query the width of band power tested.
Example: :SPECtrogram:MARKer:BWIDth?
Spectrogram Marker Audio Alert State
[:SENSe]:SPECtrogram:MARKer:ALERt:STATe
OFF|ON|0|1
[:SENSe]:SPECtrogram:MARKer:ALERt:STATe?
Use this command to turn on or off the Marker audio alert state. If the
Marker audio alert state is set to on, HSA will alert when there is signal
beyond the audio alert limit value.
Example: :SPECtrogram:MARKer:ALERt:STATe ON
Key access: Meas > Spectrum Monitor, then Marker > More 1 of 2 > Audio Alert
Vista de pagina 137